3.1 Clinical evaluation
In order to demonstrate the conformity of a medical device with the relevant general requirements and to assess undesirable side effects and the acceptability of the benefit-risk ratio, the manufacturer shall rely on clinical data providing sufficient clinical evidence. It shall take into account the characteristics and intended purpose of the medical device in order to determine and justify the level of evidence required for such demonstration. The clinical evaluation shall be planned, conducted continuously, and documented in accordance with Annex XIV, Part A of the Regulation on medical devices. The principles and procedures governing clinical evaluation are set out in Article 61 and Annex XIV, Part A of the Regulation on medical devices.
In practice, clinical evaluation is divided into two or even three phases
Phase 1: The collection and evaluation of existing data for the establishment of a clinical evaluation plan.
A first phase, conducted during the development of the medical device, involving the collection of data on known benefits and risks, the establishment of the state of the art, the assessment of the need for additional data, etc. These data include preclinical data (toxicity, mechanical and chemical testing, animal studies, etc.) and a review of the literature relating to the medical device or to an equivalent device. On the basis of the data collected, it is also necessary to determine whether a clinical investigation is required, as described below, (except for Class III medical devices for which a clinical investigation is in any case mandatory), or to establish equivalence with a device already placed on the market. This initial (preliminary) assessment thus serves as the basis for the clinical evaluation plan setting out the modalities of the second phase.
Phase 2: Implementation of the clinical evaluation plan for the assessment of the medical device’s conformity with the general requirements.
This second phase corresponds to the actual assessment of the conformity of the medical device with the general requirements referred to above, with a view to obtaining CE marking for the placing of the device on the market. Its purpose is to update and verify the quality of the data collected, to verify and demonstrate the conformity of the medical device in light of those data, and, where appropriate through the conduct of a clinical investigation, to address the various issues identified as remaining unresolved during the first phase.
| Particular attention shall be paid to the use of non-viable human or animal tissues and cells and to the associated benefit-risk ratio. In particular, in the case of species capable of transmitting TSEs, the data must make it possible to justify the use of such animal tissues and derivatives rather than other biological or synthetic substances (Article 5(2)(b) of the Regulation (EU) No 722/2012). |
The conduct of a clinical investigation may be waived (except, in principle, for Class III medical devices) when the manufacturer is able to demonstrate the equivalence of its medical device (from a technical, biological and clinical perspective), with another medical device already placed on the market. In such a case, the manufacturer may rely on the clinical data relating to the already CE-marked device in order to establish the conformity of its own medical device. (Article 61(3)(a) of the Regulation on medical devices and MDCG 2020-5: Clinical Evaluation - Equivalence A guide for manufacturers and notified bodies)
Although manufacturers of Class III medical devices are, in principle, required to conduct a clinical investigation, they may also demonstrate the equivalence of their device with other medical devices already marketed in order to support and strengthen their clinical evaluation.
Furthermore, exceptionally (except for Class III medical devices), it is possible to demonstrate conformity of a medical device with the general requirements on the basis of non-clinical data alone, provided that such an assessment, based inter alia on risk management and on technical and preclinical data, is duly justified in the technical documentation. (Article 61(10) of the Regulation on medical devices)
Finally, it is also during this second phase that the manufacturer determines and justifies the need for updates to clinical data. This conformity assessment must also serve to establish Post-Market Clinical Follow-Up (PMCF), the third phase of clinical evaluation, which falls more broadly within post-market surveillance. (Article 61(11) of the Regulation on medical devices)
Phase 3: Implementation of the Post-Market Clinical Follow-Up (PMCF)
All of these data, results, justifications and conclusions are compiled in a final report (clinical evaluation report, based on the template available here in PDF or Word format – Clinical Evaluation Assessment Report Template), which is included in the technical documentation and intended for notified bodies.
Although published prior to the adoption of the Regulations on medical devices and in vitro diagnostic medical devices, MEDDEV 2.7/1 Rev 4 - Guidelines on medical devices continues to provide manufacturers with a number of relevant indications regarding the conduct of clinical evaluation. This is supplemented by the guidance issued by the International Medical Device Regulators Forum (IMDRF), IMDRF MDCE WG/N56FINAL:2019.
3.2 Clinical investigation
What is a clinical investigation?
When the clinical data obtained through the literature review prove insufficient to establish a favourable benefit-risk ratio and/or when no equivalence can be demonstrated, the conduct of one or more clinical investigations becomes necessary.
Clinical investigation for medical devices is equivalent to clinical trials for medicines (see Clinical trials for ATMPs). It therefore consists of one or more studies conducted on human participants, with the aim of demonstrating the performance, effectiveness and safety of the medical device. The results are then incorporated into the body of clinical data supporting the clinical evaluation. As with a clinical trial for a medicinal product, a clinical investigation of a medical device must comply with a number of regulatory requirements: informed consent of participants, protection of their rights, dignity and safety, prior assessment by an ethics committee, etc. (Articles 62 to 80 and Annex XV of the Regulation on medical devices).
Clinical investigation: a mandatory requirement for Class III medical devices, subject to exceptions
As mentioned above, a clinical investigation is mandatory as part of the clinical evaluation of Class III medical devices (Article 61(2) of the Regulation on medical devices).
By way of exception, a clinical investigation is not required when 3 cumulative conditions are met:
- The device consists of a modification of a medical device already placed on the market by the same manufacturer;
- The manufacturer demonstrates that the modified device is equivalent to the already marketed device. This demonstration must be subject to approval by a notified body; and
- The clinical evaluation of the already marketed device is sufficient to demonstrate conformity of the modified device with the relevant safety and performance requirements (Article 61(4) of the Regulation on medical devices).
The exemption from the requirement to conduct a clinical investigation may also apply when the manufacturer of the modified device is not the manufacturer of the already marketed device. In addition to the previous conditions, 3 additional conditions must be met:
- “the two manufacturers have a contract in place that explicitly allows the manufacturer of the second device full access to the technical documentation on an ongoing basis”;
- “the original clinical evaluation has been performed in compliance with the requirements of [the] Regulation”; and
- the manufacturer of the second device provides the notified body with clear evidence that conditions 1 and 2 are fulfilled
(Article 61(5) of the Regulation on medical devices).
There is also an exemption from the requirement to conduct a clinical investigation for the devices referred to in Article 61(6) of the Regulation on medical devices.
These exemptions from the requirement to conduct a clinical investigation for Class III medical devices are addressed in MDCG guidance: MDCG 2023-7 Guidance on exemptions from the requirement to perform clinical investigations pursuant to Article 61(4)-(6) MDR and on sufficient levels of access’ to data needed to justify claims of equivalence.
In practice, although a clinical investigation may exceptionally not be required for Class III medical devices, it is often necessary for innovative devices, as their novelty prevents reliance on existing data relating to an equivalent device.
Submission of a clinical investigation application to the Member States concerned
The submission of a clinical investigation application by its sponsor is made to the competent national authorities (See EuroGCT | Actors and Networks). Pending the deployment of the Clinical Investigations & Performance Studies module of the EUDAMED platform, intended to provide sponsors with a single EU-wide entry point (Article 70(1) of the Regulation on medical devices), applications shall be submitted directly to the competent national authorities in accordance with national procedures.
It is then for the competent national authorities to assess the application in light of the requirements set out in Article 62 of the Regulation on medical devices.
The favourable opinion of an ethics committee in the relevant Member State is always required before a clinical investigation may commence.
However, the Regulation on medical devices distinguishes between two situations:
- Non-invasive Class IIa and IIb devices, for which validation of the application (complete dossier and clinical investigation falling within the scope of the Regulation) permits the clinical investigation to start immediately (Article 70(7)(a) of the Regulation on medical devices)
- All other medical devices, for which authorisation of the clinical investigation is required (Article 70(7)(b) of the Regulation on medical devices).
The Regulation on medical devices also allows Member States to adopt, to a certain extent, specific national rules concerning clinical investigations. The European Commission provides a list of links to the websites of competent authorities that have implemented such specific national rules.
Possibility of a single application and coordinated assessment for clinical investigations conducted in several Member States
The Regulation on medical devices also provides for the possibility for a sponsor to submit a single application for clinical investigations conducted in several Member States. In such cases, the assessment referred to in Article 78 of the Regulation on medical devices is subject to a coordinated procedure between the Member States concerned, one of which is designated as the coordinating Member State.
As the single application is intended to be submitted via the EUDAMED platform, its effective implementation must await the deployment of the Clinical Investigations & Performance Studies module.
However, in February 2025, the European Commission launched a call for applications to participate in a Pilot coordinated assessment of clinical investigations and performance studies.
“This pilot will allow sponsors to submit a single application for pilot coordinated assessments, assuring more harmonised interaction with the Member States approving the clinical investigation or performance studies. Additionally, participant sponsors will help the competent authorities of Member States with building a fast and fit for purpose EU system for coordinated assessment once its mandatory operation starts as required”.
18 EU countries (Austria, Belgium, Czechia, , Estonia, Finland, France, Germany, Greece, Ireland, Latvia, Luxembourg, Poland, Portugal, Romania, Slovenia, Spain, Sweden and The Netherlands) and Norway are participating.
Recording and reporting of adverse events occurring during clinical investigations
During the clinical investigation, the sponsor shall record any adverse event or device deficiency occurring with the investigational device and shall notify the relevant Member State(s) in the event that the occurrence is serious (Article 80 of the Regulation on medical devices). Here again, notification shall be made directly to the competent national authorities pending the deployment of the Clinical Investigations & Performance Studies module of the EUDAMED platform.